Share With A Friend:

ReadMyHeart Specifications

Specifications:
Input impedance > 10M ohm
Input dynamic range +/- 2 mV
Bandwidth 0.15 - 40 Hz
CMRR (Common Mode Rejection Ratio) > 60 db
A/D conversion 12 bits
Sampling frequency 250 samples/s
Memory 15 seconds of reading, 30 recordings
Display LCD display panel
Input Dry conduction electrodes and/or external auxiliary electrodes
Please note: The software we provide is not compatible with Macintosh
Power Supply 1.5V (AAA) x 2
Size 124 x 78 x 22 mm
Weight 150g (Not including batteries)
Measurement Range:
Average heart rate 45 to 180 bpm
ST value -3 to +3 mm
QRS interval < 0.20 sec
Storage Temperature -4 to 122 F (-20 to 50C)
Operating Temperature 50 to 104 F (10 to 40C)
Humidity 25% to 95%